Ames Research Center; Center for Gravitational Biology Research
Home Contacts Links About Us    Facilities      Questionnaire      Publications  


1-FOOT DIAMETER CENTRIFUGE
divider

Operational Characteristics:
  • Acute or chronic exposures to 180 G
  • Cell, tissue, and small organism payloads
  • Controlled atmosphere
The 1-Foot Diameter Centrifuge is designed to test the effects of hypergravity on cultured cells and small organisms. This system consists of a standard table top centrifuge equipped with a GH-3.8 rotor. The rotor has platforms that swing out to a maximum radius of 6.4 inches. The centrifuge has been modified to control speeds from 45-1000 rpm (1.4-180 X Earth Gravity) and to circulate air from the chamber of an adjacent water-jacketed tissue culture incubator. The incubator is used to control temperature, CO2, O2, and humidity in the centrifuge. The incubator also can be bypassed to provide an ambient environment in the centrifuge. An identical incubator is located nearby for stationary controls.

The temperature, CO2, and relative humidity in the centrifuge are monitored using sensors and a remote data acquisition system. The sample carriers consist of 4 microtitre plate platforms, each measuring 5.25 in. X 3.5 in. The carriers accommodate 96-well microtitre dishes or other dishes of similar size. The sample mass cannot exceed 215 grams on each of the platforms.
centrifuge


 Performance limits and specifications:

 Radius: 6.4 in
 Payload: 1.9 lbs
 Max G: 180 G
 Max RPM: 1000



Rollover graphic Rollover graphic Rollover graphic Rollover graphic Rollover graphic Rollover graphic Rollover graphic Rollover graphic